Standardization and Validation of Non-invasive Monitoring of End Tidal Carbon Dioxide in Neonates via Nasal Cannula: An Observational Study

Document Type : Original Article

Authors

1 Department of Respiratory Therapy, School of Allied Health Sciences, Manipal Academy of Higher Education, Manipal, Karnataka, India

2 Head of Neonatal Intensive Care Unit, Kasturba Hospital Manipal, Karnataka, India

Abstract

Background: Carbon dioxide (CO2) is a by-product of cellular metabolism, which could be considered as a reflection of metabolism, circulation, and ventilation. Arterial blood gas analysis (ABG) is the gold standard of monitoring for CO2. However, is an expensive method leading to blood loss and iatrogenic anemia. In addition, each sample is only a snapshot view of the sampling moment. End-tidal CO2 (ETCO2) measurement gives a non-invasive and continuous monitoring of exhaled CO2. Therefore, this study aimed to validate the non-invasive CO2 measurement performed by nasal cannula and evaluate the correlation with partial pressure of arterial CO2 (PaCO2 ) in neonates.
Methods: This single-center observational study was conducted in the Neonatal Intensive Care Unit (NICU), Kasturba Hospital, Manipal, India. PaCO2 was reported on routine ABG within a 15-minute interval of ABG sampling. Moreover, partial pressure of end-tidal carbon dioxide (PetCO2) was noted at continuous 30-sec intervals (i.e., 30, 60, 90, ..., 180) up to 3 minutes. The values of PaCO2 and PetCO2 were found to be correlated with Pearson correlation and were shown by scattered plot. Regression analysis was used to get the prediction equation and the variance.
Results: A total of 70 samples were taken in the initial phase to study the correlation between PaCO2 and measured PetCO2. Pearson correlation showed a moderate positive correlation (r= 0.589) between PetCO2 and measured PaCO2. Regression analysis demonstrated a variance of 33.8% between the measured PetCO2 and PaCO2, which was statistically significant (P<0.001). A prediction equation was obtained for PaCO2. In the final phase, 20 samples were recruited to standardize and validate the prediction equation. The PaCO2 was calculated using the predicted equation and a new prediction equation was obtained.
Conclusion: According to the findings of this study, there is a good correlation (r=0.681) between the non-invasively measured PetCO2 and PaCO2.

Keywords


1. Bhende MS. End-tidal carbon dioxide monitoring in pediatrics-clinical applications. J Postgrad Med. 2001; 47(3):215-8.
 2. Sanders AB. Capnometry in emergency medicine. Ann Emerg Med. 1989; 18(12):1287-90.
3. Bhat YR, Abhishek N. Mainstream end-tidal carbon dioxide monitoring in ventilated neonates. Singapore Med J. 2008; 49(3):199-203.
4. Pichler G, Urlesberger B, Schmölzer G, Müller W. Effect of tilting on cerebral haemodynamics in preterm infants with periventricular leucencephalomalacia. Acta Paediatr Int. 2004; 93(1):70-5.
5. Graziani LJ, Spitzer AR, Mitchell DG, Merton DA, Stanley C, Robinson N, et al. Mechanical ventilation in preterm infants: neurosonographic and developmental studies. Pediatrics. 1992; 90(4):515-22.
6. Epstein MF, Cohen AR, Feldman HA, Raemer DB. Estimation of PaCO2 by two noninvasive methods in the critically ill newborn infant. J Pediatr. 1985; 106(2):282-6.
 7. Kugelman A, Zeiger-Aginsky D, Bader D, Shoris I, Riskin A. A novel method of distal end-tidal CO2 capnography in intubated infants : comparison with arterial CO2 and with proximal mainstream end-tidal CO2. Pediatrics. 2008; 122(6):e1219-24.
8. Jin Z, Yang M, Lin R, Huang W, Wang J, Hu Z, et al. Application of end-tidal carbon dioxide monitoring via distal gas samples in ventilated neonates. Pediatr Neonatol. 2017; 58(4):370-5.
 9. Bowe EA, Boysen PG, Broome JA, Klein EF Jr. Accurate determination of end-tidal carbon dioxide during administration of oxygen by nasal cannulae. J Clin Monit. 1989; 5(2):105-10.
 10. Liu SY, Lee TS, Bongard F. Accuracy of capnography in nonintubated surgical patients. Chest. 1992; 102(5):1512-5.
11. Tai CC, Lu FL, Chen PC, Jeng SF, Chou HC, Chen CY, et al. Noninvasive capnometry for end-tidal carbon dioxide monitoring via nasal cannula in nonintubated neonates. Pediatr Neonatol. 2010; 51(6):330-5.
 12. Moses JM, Alexander JL, Agus MS. The correlation and level of agreement between end-tidal and blood gas pCO2 in children with respiratory distress: a retrospective analysis. BMC Pediatr. 2009; 9(1):20.
 13. Coates BM, Chaize R, Goodman DM, Rozenfeld RA. Performance of capnometry in non-intubated infants in the pediatric intensive care unit. BMC Pediatr. 2014; 14:163.